Description
The Infant Incubator YG0426 offers advanced features for the precise care of premature and critically ill newborns. Equipped with a 7-inch LCD touch screen control panel, the YG0426 allows healthcare providers to easily monitor and adjust critical parameters such as air and skin temperature. This incubator provides servo-controlled air and skin temperature regulation, ensuring the newborn is maintained within a safe, stable environment. The air mode temperature ranges from 25.0°C to 37.0°C, while the skin mode temperature operates between 34.0°C and 37.0°C, offering optimal conditions for neonatal growth and recovery.
The YG0426 also features a variety of safety alarms, including high-temperature mode alarms and skin and air mode temperature alarms, to notify caregivers if the temperature goes beyond safe levels. Additional conveniences include free-step mattress tilting adjustment, a removable humidity trough for easy cleaning, and the inclusion of a transfusion shelf and tray for medical procedures. Optional features like a phototherapy unit, electric weighing function, and an electrical lift make the YG0426 a versatile and adaptable incubator, providing all-around care for neonates in various medical conditions.
Market Price
The Infant Incubator YG0426 is available in the market at a price range of approximately $9.949 to $10.000, depending on the configuration and selected options, such as the phototherapy unit, electric weighing function, and electrical lift. This price reflects the high-end technology, precision control, and customizability that make the YG0426 an ideal choice for neonatal care units in hospitals, clinics, and specialized healthcare facilities.
Frequently Asked Questions
- What is the temperature control range of the YG0426?
The YG0426 offers air mode temperature control from 25.0°C to 37.0°C, and skin mode temperature from 34.0°C to 37.0°C, providing optimal warmth for newborns. - Does the incubator include safety alarms?
Yes, the YG0426 is equipped with various alarms, including high-temperature mode alarms, air mode alarms, and skin mode alarms to ensure the newborn is always in a safe environment. - What optional features are available with the YG0426?
The YG0426 can be equipped with a phototherapy unit for jaundice treatment, an electric weighing function for monitoring infant growth, and an electrical lift for easy adjustment.
Advantages and Disadvantages
- Advantages:
Precise Control: With its temperature and humidity control capabilities, the YG0426 ensures the ideal environment for neonates.
Safety Alarms: The incubator features multiple alarms to alert caregivers of any temperature deviations, ensuring peace of mind.
Customizable Options: Optional accessories such as phototherapy, electric weighing function, and an electrical lift enhance the incubator’s versatility.
Easy Maintenance: The removable humidity trough simplifies cleaning and maintenance, keeping the incubator in optimal condition.
- Disadvantages:
Higher Initial Cost: The YG0426 may have a higher initial cost compared to basic incubators, which might be a consideration for smaller healthcare facilities.
Size Requirements: Due to its features, the incubator requires ample space, which may not be ideal for all care environments.
Use of the Product in the Field
The Infant Incubator YG0426 is an essential tool in Neonatal Intensive Care Units (NICUs) and hospitals where newborns, especially premature or ill infants, require a controlled environment. Its ability to provide precise temperature, humidity, and optional phototherapy makes it ideal for supporting neonatal growth and development. The incubator’s versatility, including the optional electric lift and electric weighing function, allows healthcare providers to perform critical medical procedures with ease, ensuring comprehensive care.
Recommendations
To ensure the YG0426 continues to function at peak performance, regularly clean the humidity trough and check the alarms and temperature settings. Periodically verify the accuracy of the temperature and humidity sensors to maintain precise control. Additionally, use the phototherapy unit when needed for jaundice treatment to enhance neonatal recovery. Consider using the electric weighing function to monitor infant progress and ensure proper growth during their stay.
Technical Specifications
Model | YG0426 |
Electrical requirement | ~220V 50Hz |
Heater Power | 550VA |
Air mode temperature range | 25.0°C-37.0°C |
Skin mode temperature range | 34.0°C-37.0°C |
Air mode temperature alarm | <38.0°C |
Skin mode temperature alarm | < 40.0°C(Air temperature) |
Air mode high temperature range | 37.1°C~39.0°C |
Skin mode high temperature range | 37.1°C~38.0°C |
Hight temperature mode alarm | < 40.0°C (Air temperature) |
Air mode temperature display range | 20.0°C-45.0°C |
Skin mode temperature display range | 25.0°C~45.0°C |
Air mode alarm | ±3.0°C |
Skin mode alarm | ±1.0°C |
Skin temperature sensor accuracy | ±0.3°C |
Temperature control accuracy | ≤0.5°C |
Mattress temperature uniformity | ≤0.8°C |
Humidity control range | 30-90%RH |
Humidity display range | 0-99.9%RH |